• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC SYMPHONY OCT SYSTEM AO HANDLE, SMALL WITH SWIVEL; MISC ORTHO SURGICAL INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC SYMPHONY OCT SYSTEM AO HANDLE, SMALL WITH SWIVEL; MISC ORTHO SURGICAL INSTR Back to Search Results
Catalog Number 202000500
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: the manufacturing date is currently not available.Therefore, the full udi is currently not available.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that before surgery on (b)(6) 2024, when the ao handles in question were cleaned and sterilized in the central supply room, a blood-like dirt was found on them, and their use was stopped.Another handle was used and the surgery was completed successfully with no delay.The patient outcome was reported to be stable.This report involves one symphony oct system ao handle, small with swivel this is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: an analysis of the product could not be performed since a physical sample was not received for evaluation.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.Device history review: a review of the receiving inspection (ri) for ao handle small w swivel was conducted identifying that lot number km898250 released in two batches.Batch1: lot qty of (b)(4) units were released jul 30, 2020 with no discrepancies.Batch2: lot qty of (b)(4) units were released jul 30, 2020 with no discrepancies.Supplier: (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMPHONY OCT SYSTEM AO HANDLE, SMALL WITH SWIVEL
Common Device Name
MISC ORTHO SURGICAL INSTR
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key20042548
Report Number1526439-2024-02111
Device Sequence Number7977614
Product Code LXH
UDI-Device Identifier10705034532622
UDI-Public(01)10705034532622(10)KM898250
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number202000500
Device Lot NumberKM898250
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/25/2024
Supplement Date Received by Manufacturer10/01/2024
Initial Report FDA Received Date08/22/2024
Supplement Report FDA Received Date10/02/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/30/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AO HANDLE MEDIUM W RATCHET.
Patient SexUnknown
-
-