• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTFLOW, INC. HEARTFLOW ANALYSIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTFLOW, INC. HEARTFLOW ANALYSIS Back to Search Results
Model Number 3.33.0.4
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/06/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of heartflow's internal review, we identified a potential computed tomography (ct) dataset was incorrectly accepted for processing the heartflow analysis.Hfid# bjh-24qlf6-jqbh: the investigation identified that the available ct dataset did not meet heartflow's image quality requirements.Heartflow incorrectly assessed image quality in this dataset during the acceptance of this case due to misinterpretation of the ct data by the automated technology and inspection process.The accuracy of this heartflow analysis cannot be confirmed due to the decreased confidence in modeling through artifact-affected regions.The customer was notified of the investigation results and they did not indicate a safety event with the patient.They were accepting of the findings and appreciated heartflow's commitment to quality control.Heartflow's analysis instructions for use include the following warnings: due to the variability in the heartflow analysis, the ffrct values should be reviewed as one of several clinical data points to be used in conjunction with the patient's original ct images, clinical history, symptoms, and other diagnostic tests, as well as an appropriately trained clinician's clinical judgment, to evaluate the patient.The heartflow analysis process is dependent on the quality of the imaging data provided.The ffrct values may be affected by assumptions needed to resolve anatomy in areas of uncertainty, whether provided by the physician or made by heartflow case analysts.
 
Event or Problem Description
Heartflow identified a potential computed tomography (ct) dataset was incorrectly accepted for processing an analysis.
 
Additional Manufacturer Narrative
Corrected data: h.4 -device manufacture date was inadvertently left blank in the initial report.The correct date is 06-aug-2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTFLOW ANALYSIS
Common Device Name
HEARTFLOW ANALYSIS
Manufacturer (Section D)
HEARTFLOW, INC.
331 e evelyn ave
mountain view CA 94041
Manufacturer (Section G)
HEARTFLOW, INC.
331 e evelyn ave.
mountain view CA 94041
Manufacturer Contact
avelina alvarado
331 e evelyn ave.
mountain view, CA 94041
6502411221
MDR Report Key20043701
Report Number3021637148-2024-00007
Device Sequence Number7955114
Product Code PJA
UDI-Device Identifier00853341006060
UDI-Public(01)00853341006060(10)FFRCT_3.33.0.4(11)2024-08-06
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K213857
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3.33.0.4
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/22/2024
Supplement Report FDA Received Date09/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/06/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-