EXACTECH, INC. LOGIC FEMORAL PS CEM LEFT SZ 2.5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-010-01-0225 |
| Medical Device Problem Codes |
Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Loss of Range of Motion (2032); Osteolysis (2377); Joint Laxity (4526); Implant Pain (4561)
|
| Date of Event |
09/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported via legal documentation that approximately 78 months after a left total knee replacement procedure, the patient underwent a revision procedure to address polyethylene wear, pain, joint laxity, pitting of the tibial insert, polyethylene wear of the patellar component, osteolysis of the tibia and femur, loose femoral component, and limited range of motion.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and femoral loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.D4: corrected d10: concomitant devices: 2582643, 02-010-01-0225 - logic femoral ps cem left sz 2.5.4271380, 200-02-35 - three peg patella 35mm.G4: corrected.H6: corrected medical device and investigation clinical codes.
|
| |
|