| Catalog Number |
21-4475-24 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter occupation unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the sheath's outer surface (the sky-blue part) was split in half; normally, the silicon valve (transparent round valve) was left alone, not split in half.There was patient involvement, and no patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6.Evaluation codes: updated.At the time of this investigation the physical device was not received at the investigation site.One (1) photo and one (1) video were provided for evaluation.In the photo provided it is observed that the sheath is being split during use and the valve doesn¿t split.On video it is shown that the sheath is being split and the user tried multiples times to break apart the valve until some tweezers are used to separate it.According to the video, the complaint was confirmed, however the root cause was not determined.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.If the product is returned the manufacturer will re-open the complaint for further device analysis.
|
| |
|
Search Alerts/Recalls
|