| Catalog Number |
A319516AM |
| Medical Device Problem Codes |
Short Fill (1575); Component Misassembled (4004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that operating room has found 3 foley trays were the 10cc syringe of sterile water had no sterile water in it.They had one label, but it does not have a lot # on it.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid as requested.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that operating room has found 3 foley trays were the 10cc syringe of sterile water had no sterile water in it.They had one label, but it does not have a lot # on it.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was available for evaluation and further investigation did not result in any additional findings.Root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be tip cover came off during shipping or in processing.The dhr review could not be performed without a lot number.A labeling review was not performed because labeling could not have prevented the reported failure.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|