MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. GASTROINTESTINAL VIDEOACOPE; GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. GASTROINTESTINAL VIDEOACOPE; GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-LV1
Medical Device Problem Codes Poor Quality Image (1408); Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was observed, during the device inspection.That the gastrointestinal videoscope exhibited foreign material in the upward/downward and right/left control knob.There were no reports of patient involvement.
 
Event or Problem Description
It was reported the subject device had an image issue.On september 2, 2024 the customer confirmed that the image quality was not good and the image was blurred.The issue occurred during maintenance.There were no reports of patient harm.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Additional information added to fields b5, d5, e1, g2, h3 and h6.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.The device was returned to olympus for inspection, and the customer's complaint of blurry image was not confirmed.Based on the results of the investigation, a probable cause was not identified.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: "chapter 3 preparation and inspection.The equipment prepared before using this endoscope and procedures for inspection of the endoscope and equipment are described in this chapter.3.1 the workflow of preparation and inspection.The workflow of preparation and inspection is shown below.Before each case, prepare and inspect this endoscope as instructed below.Inspect other equipment to be used with this endoscope as instructed in their respective instruction manuals.Should any irregularity be observed after inspection, follow the instructions as described in chapter 5, ¿troubleshooting¿.If the endoscope malfunctions, do not use it.Return it to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Warning: never use the endoscope on a patient if any irregularity is observed.The irregular endoscope may compromise patient or user safety and may result in more severe equipment damage.In addition, it may pose an infection control risk."."chapter 5 troubleshooting.The countermeasures against troubles are described in this chapter.5.1 troubleshooting.If any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2, ¿troubleshooting guide¿.If the problem still cannot be resolved, send the endoscope to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Also, should any irregularity be observed while using the endoscope, stop using it immediately and withdraw the endoscope from the patient as described in section 5.3, ¿withdrawal of the endoscope with an irregularity¿.Warning: never use the endoscope on a patient if any irregularity is observed.The irregular endoscope may compromise patient or user safety and may result in more severe equipment damage.In addition, it may pose an infection control risk".Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GASTROINTESTINAL VIDEOACOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20050243
Report Number9610595-2024-16845
Device Sequence Number8117498
Product Code FDS
UDI-Device Identifier04953170312373
UDI-Public04953170312373
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K111756
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGIF-LV1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2024
Supplement Date Received by Manufacturer09/27/2024
Initial Report FDA Received Date08/23/2024
Supplement Report FDA Received Date10/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-