Model Number 70914-70
Medical Device Problem Codes Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
Health Effect - Clinical Codes Hyperglycemia (1905); Loss of consciousness (2418); Convulsion/Seizure (4406)
Date of Event 06/21/2024
Type of Reportable Event Serious Injury
Event or Problem Description
An error message was reported with the abbott diabetes care (adc) device.The customer received an unspecified error message on their meter display and was unable to obtain glucose readings.The customer experienced a seizure and a loss of consciousness and was able to self-treat with metformin which was provided to them by a healthcare professional.There was no report of death or permanent impairment associated with this event.
 
Additional Manufacturer Narrative
The product has been requested back for investigation and the customer agreed to return the device.A follow-up report will be submitted upon completion of investigation activities or if additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
An error message was reported with the abbott diabetes care (adc) device.The customer received an unspecified error message on their meter display and was unable to obtain glucose readings.The customer experienced a seizure and a loss of consciousness and was able to self-treat with metformin which was provided to them by a healthcare professional.There was no report of death or permanent impairment associated with this event.
 
Additional Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time, the product has not yet been returned for this complaint.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs (device history review) for the freestyle freedom lite meter and freestyle strip was reviewed and the dhrs showed the freestyle freedom lite meter and freestyle strip meet specifications prior to release to distribution.Retain testing was performed for freestyle strip and all units performed in specification and passed.If product is returned, a physical investigation will be performed per adc's established processes and procedures and a report will be submitted upon completion of investigation.This also serves as a correction report section h4 ( device mfg date) was incorrectly updated in the previous report.Correction has been made all pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE FREEDOM LITE
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key20050318
Report Number2954323-2024-32062
Device Sequence Number7934097
Product Code NBW
UDI-Device Identifier00699073714992
UDI-Public(01)00699073714992(10)01A020D(17)311221
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K092602
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/31/2026
Device Model Number70914-70
Device Lot Number1230222
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2024
Supplement Date Received by Manufacturer01/12/2025
Initial Report FDA Received Date08/23/2024
Supplement Report FDA Received Date01/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-