| Model Number |
AA07 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Repair request initiated for device with the report of difficulty with insertion.No patient impact reported.Repair escalated to product event on evaluation due to detached bearing.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis :evaluation determined that distal bearing are detached.The likely cause of failure was identified as inadequate preventive maintenance.It was also noted that the burr cannot be inserted into the attachment and laser markings are faded.Event notified date:21-jun-2024 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|