Issue identified during product analysis.Medtronic conducted an investigation based upon all information received.It was reported that these 980 ventilators failed the extended self test (est) with an "inspiratory auto zero solenoid not operational" message.The device was available for evaluation.The service personnel (sp) inspected the ventilator, confirmed the reported issue and replaced the inspiratory flow module (ifm) printed circuit board assembly (pcba).The ventilator passed all required calibrations and tests per manufacturer specifications at the time of service.One ifm pcba was returned for failure investigation.The part was visually inspected and no anomalies were observed.The part was attached to the test ventilator and powered up but failed the extended self test (est) circuit pressure test ¿inspiratory autozero solenoid not operational¿.After a thorough investigation, a faulty so1 in the ifm pcba was identified.Analysis also determined the ifm pcba was prior to the implementation of a change to address autozero solenoid (so1) failures.If an so1 failure were to occur while in use on a patient the ventilator response would be to enter backup ventilation (buv).The cause of the event was identified as a faulty auto zero solenoid (so1) on the ifm pcba.There is an existing previous internal i nvestigation regarding this issue.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|