| Catalog Number |
0233050300 |
| Medical Device Problem Codes |
Failure to Shut Off (2939); Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a thermal event.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was previously closed based on product not received; however, the product has now been physically received at stryker endoscopy, usa and the investigation has been re-opened.Investigation as follows is now based on product received.Customer has a bad safelight cable.S/n: (b)(6), stays illuminated even when adapter is disconnected.The failure(s) identified in the investigation is consistent with the complaint record.Probable root causes are: bad reed switch the product was returned for investigation and the failure mode will be monitored for future reoccurrence.The manufacturer date in not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a thermal event.
|
| |
|