| Model Number |
1128-001102-01-66 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during incoming inspection, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing including on/off current testing without duplicating the report.The customer received a replacement device.The battery used was not returned as part of this investigation.No trend is associated with reports of this type.
|
| |
|
Search Alerts/Recalls
|