• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG. ASSY. PEEK SOFT EYELET RC ANCHOR 5.5MM WITH NEEDLES; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG. ASSY. PEEK SOFT EYELET RC ANCHOR 5.5MM WITH NEEDLES; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 3910-400-015
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the anchor broke during the procedure.All pieces were retrieved.
 
Event or Problem Description
It was reported that the anchor broke during the procedure.All pieces were retrieved.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: intraline broke.Probable root cause: process: sutures not assembled to specification.Handle not manufactured to specifications.The reported failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG. ASSY. PEEK SOFT EYELET RC ANCHOR 5.5MM WITH NEEDLES
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key20055547
Report Number0002936485-2024-00597
Device Sequence Number8117714
Product Code HWC
UDI-Device Identifier07613154644575
UDI-Public07613154644575
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K071157
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3910-400-015
Device Lot Number23259AE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/26/2024
Supplement Date Received by Manufacturer07/26/2024
Initial Report FDA Received Date08/23/2024
Supplement Report FDA Received Date11/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-