| Model Number |
AVA10 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3:product analysis: evaluation determined that distal bearing are detached.The likely cause of failure is inadequate preventive ma intenance.It was also noted that laser markings are faded and color rings are faded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with the report of seized/not running.It was reported there was no patient involvement. repair escalated to a product event on evaluation due to the bearing detachment.
|
| |
|
Search Alerts/Recalls
|