| Model Number |
GIF-N180 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed during the device inspection, that the probe cover of the videoscope had foreign material.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: h3, h6 and h11.This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation foreign material in the device could not be identified, there was no physical damage where the foreign material was found, and there were no reported reprocessing deviations from the ifu.It is possible the foreign material was not removed during reprocessing due to damage on guide pin of connector.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) section below: ifu states the detection method in evis exera ii gif/cf/pcf type 180 series operation manual chapter 3 preparation and inspection.Ifu states the preventive measure in evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).Olympus will continue to monitor field performance for this device.
|
| |
|