| Model Number |
610-2221411-01 |
| Medical Device Problem Codes |
Pacing Problem (1439); Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during training by facility staff, the device was not able to pace and would not provide a sync shock.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
A zoll medical representative went onsite to evaluate the device and the customer's report was not replicated or confirmed.The device was put through extensive functional testing including pace and sync shock using a simulator without duplicating the report.The device was recertified and returned to service.No trend is associated with reports of this type.
|
| |
|