The device was not returned to stryker sustainability solutions for evaluation as multiple attempts were made but unsuccessful.As the complaint device was not returned, a conclusive root cause could not be determined and the reported event could not be confirmed.A review of the dhr for the reported lot/serial number supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to: not plugged in completely to recognize bar code or plugged into incompatible console ancillary equipment failure device plugged into multiple generators.The instructions for use (ifu) state: open the pouch, remove the device and the card from the pouch.Electric shock hazard ¿ do not connect wet accessories to the ligasure system.Position instrument cords to avoid contact with the patient or other cords.Do not wrap cords around metal objects.This may induce currents that could lead to shocks, fires, or injury to the patient or surgical team.Examine all ligasure system and instrument connections before using.Improper connections may result in arcing, sparks, accessory malfunction, or unintended surgical effects.- inspect the instrument and cords for breaks, cracks, nicks, or other damage before use.Failure to observe this caution may result in injury or electrical shock to the patient or surgical team or cause damage to the instrument.If damaged, do not use.Confirm proper ligasure system settings before proceeding with surgery.Connect adaptors and accessories to the electrosurgical unit only when the unit is off or in standby mode.Failure to do so may result in injury or electrical shock to the patient or operating personnel.Use caution when handling the instrument between uses to avoid accidental activation of the ligasure system.Do not place the instrument on the patient or drapes when not in use.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
|