ALCON RESEARCH, LLC - HOUSTON CONSTELLATION VISION SYSTEM, ACCESSORY, IRRIGATION/ASPIRATION TUBING SET; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750918
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that the tube was blocked and could not be aspirated normally during cataract/vitrectomy combination surgery.The surgery was completed after replacement of the irrigation/aspiration tubing set.There were no other abnormalities.The patient returned to the ward smoothly after completing the surgery.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned sample was visually inspected, and no obvious defects were found.A calibrated system console representing the current software version was used to test the sample.The sample could prime and pass intraocular pressure (iop) calibration successfully.No anomalies were observed during priming.No system message code was generated during testing.Fluid flowed from the balanced salt solution (bss) bottle to the drain bag without any interfering.No leakage was detected.Cleaning process was able to be performed after functional test had completed.The root cause of the customer's complaint could not be established; the returned sample functioned per specifications.After an investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM, ACCESSORY, IRRIGATION/ASPIRATION TUBING SET
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20064491
Report Number1644019-2024-01922
Device Sequence Number15577430
Product Code HQC
UDI-Device Identifier00380657509188
UDI-Public(01)00380657509188(17)250430(11)230502(10)15KK4N
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K101285
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Catalogue Number8065750918
Device Lot Number15KK4N
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/31/2024
Supplement Date Received by Manufacturer11/22/2024
Initial Report FDA Received Date08/26/2024
Supplement Report FDA Received Date12/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-