| Model Number |
LF-GP |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed, during the device inspection.That the fiberscope exhibited the coating peeled on the connecting tube.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 15 years since the subject device was manufactured.Based on the results of the investigation, olympus confirmed the suggested event, and it was presumed that the event was caused by stress of repeated use, external factors, or handling of the device.However, a definitive root cause could not be identified.The event can be detected and prevented by following the instructions for use: it was confirmed that instructions for the tracheal intubation fiberscope lf-tp/dp/gp chapter 3, ¿preparation and inspection¿ section 3.2, ¿preparation and inspection of the endoscope¿ describes how to inspect for the subject event.Olympus will continue to monitor field performance for this device.
|
| |
|