Olympus reviewed the following literature titled "association of local steroid injection as a risk factor for electrocoagulation syndrome after esophageal endoscopic submucosal dissection." this study aimed to investigate the risk factors associated with postendoscopic submucosal dissection electrocoagulation syndrome (peecs) in patients undergoing esophageal endoscopic submucosal dissection (esd), focusing on the significance of local steroid injections.In a total of 202 consecutive patients, peecs was recorded in 98 patients.Patients with peecs exhibited a larger tumor size (25.0 vs.17.0 mm, p=0.002), longer procedure (40.0 vs.29.5 min, p=0.021) and hemostasis times (5.0 vs.3.5 min, p=0.004), required greater submucosal injection volume (60.0 ml vs.50.0 ml, p=0.030), and had a lower rate of local steroid injection (4.1% vs.12.5%, p=0.029) than those without peecs.Type of adverse events/number of patients postendoscopic submucosal dissection electrocoagulation syndrome (peecs) (98 patients) (include as follows): fever (91 patients) leukocytosis (91 patients) chest pain (82 patients) delayed bleeding (3 patients) extended-spectrum--lactamase negative escherichia coli positive blood culture (1 patient) esophageal peecs was defined by satisfying at least two of the following criteria: fever =37.8° c, leukocytosis = 10,800/mm3, and localized chest pain = 5/10 points as associated on a numeric rating scale within 24 hours after esd.Among the 98 patients with peecs, 93 (94.9%) were administered empirical antibiotics after blood and culture tests.In one patient, the blood culture was positive for extended-spectrum--lactamase negative escherichia coli.The patient was treated with intravenous antibiotics without complications.Similarly, in this study, only delayed bleeding was observed in three (1.5%) patients.Treatment/intervention information for delayed bleeding was not provided in the literature.
|
|
This report is related to the following linked patient identifiers: (b)(6).The device was not returned.Attempts were performed to obtain additional information, but no response was received.Should additional relevant information become available, a supplemental report will be submitted.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial and/or lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the relationship between the device and the adverse event cannot be confirmed.There was no complaint reported on the subject device.There is no evidence of an olympus device malfunction.Therefore, the root cause cannot be determined.This supplemental report includes additional information received from the author.B5 updated accordingly.Also, a correction has been made to d4 (udi).Since the device was not returned to olympus for evaluation, and the serial/lot number is unknown, the udi is unknown at this time.Olympus will continue to monitor field performance for this device.
|