Model Number M0061801550
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during preparation prior to ureteroscopy procedure, the contour vl stent was noticed to be broken into two pieces along the shaft.Another of the same stent was used to successfully complete the procedure.No patient complications were reported.The condition of the patient at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of stent shaft break.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of stent shaft break.Block h11: the contour vl ureteral stent has been returned.During the visual and magnification analysis, it was observed that the stent shaft was detached.The reported event of stent shaft break was confirmed.It is possible to conclude that this problem could be caused by operational factors, the retrieval line may be removed before placement at the physician's discretion.If the retrieval line gets entangle in the shaft and is removed using excess force, the stent can get detached during the preparation affecting the performance of the device.Therefore, all compiled information on this investigation determines that the most probable cause is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.
 
Event or Problem Description
It was reported that during preparation prior to ureteroscopy procedure, the contour vl stent was noticed to be broken into two pieces along the shaft.Another of the same stent was used to successfully complete the procedure.No patient complications were reported.The condition of the patient at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR VL
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key20068656
Report Number2124215-2024-53100
Device Sequence Number8318092
Product Code FAD
UDI-Device Identifier08714729058410
UDI-Public08714729058410
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K974541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/20/2025
Device Model NumberM0061801550
Device Catalogue Number180-155
Device Lot Number0030574545
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2024
Supplement Date Received by Manufacturer10/03/2024
Initial Report FDA Received Date08/26/2024
Supplement Report FDA Received Date10/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-