Catalog Number 222229
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
This is report 2 of 2 for (b)(4).It was reported from china that during a rotator cuff repair procedure, it was discovered that the anchor on the 4.5 healix br anchor w/ocord device had broken off.All the broken parts were removed, and another device was used to continue the surgery.However, the same issue occurred with the replacement device.It was not reported if there were any delays to the surgical procedure.It was reported that a spare device was used to complete the surgery.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2024.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.E1: the reporter¿s phone number was not provided.Device history record review: a device history record review was performed which indicated that there was a non-conformance unrelated to the reported malfunction.All the issues identified were resolved prior to product release.Investigation summary the product has not returned to depuy synthes mitek, however photos were provided for review.The photo investigation revealed that 4.5 healix br anchor w/ocord anchor was broken at the distal end.No other anomalies could be observed.The overall complaint was confirmed as the observed condition of 4.5 healix br anchor w/ocord would contribute to the complained device issue.Based on the investigation findings, the potential cause for the broken anchor can be traced to off axis insertion and levering during insertion.As per ifu, inserting the awl less than the specified depth, axial misalignment or levering with the anchor upon insertion, may result in anchor fracture, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5 HEALIX BR ANCHOR W/OCORD
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key20069729
Report Number1221934-2024-02930
Device Sequence Number20779562
Product Code MAI
UDI-Device Identifier10886705007769
UDI-Public(01)10886705007769(10)250L701(17)261231
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number222229
Device Lot Number250L701
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2024
Initial Report FDA Received Date08/26/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-