| Catalog Number |
72204397 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an acl procedure, the screw broke upon insertion into the acl tunnel.All the broken pieces of the screw were successfully removed from the patient and a back up screw of the same size was used to complete the procedure.No surgical delay and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: additional information on: b5, b7, d4 & h4.
|
| |
|
Event or Problem Description
|
|
It was reported that during an acl procedure, the biosure regenesorb screw broke upon insertion into the acl tunnel.All the broken pieces of the screw were successfully removed from the patient with the help of gillies forceps, and a back up screw of the same size was used to complete the procedure, no void was left in the patient.No delay and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor material specification, found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|