OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, CYLINDER, 12° AND 30°, STERILE, SINGLE USE, 12 PCS.; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, MONOPOLAR, SINGLE-USE
|
Back to Search Results |
|
| Model Number |
A22266C |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, the resection electrode tip broke while being ignited and cylinder came off.The cylinder was retrieved.The issue occurred during a therapeutic transurethral resection of the prostate (turp) procedure.There were no reports of patient harm.The intended procedure was completed with a similar device.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.The device was returned to olympus for evaluation and event was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over one (1) year since the subject device was manufactured.Based on the results of the investigation, the distal end of the electrode was disconnected and showed signs of charring.Olympus was unable to perform a functional inspection as the one side of the insulation was detached from the electrode.The wire at the distal end melted away to the fork insulation.The malfunction likely occurred due to wear and tear.However, a definite root cause could not be determined.Additionally, it should be mentioned that the instructions for instructions for use (ifu) note the loop at the distal end of the electrode wears out during use and may break, burn or melt.Olympus will continue to monitor the field performance of this device.
|
| |
|