|
EXACTECH, INC. LOGIC TIBIA PS MOD INSRT SZ 2.5 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-35-2509 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Cyst(s) (1800); Pain (1994); Swelling/ Edema (4577)
|
| Date of Event |
06/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant devices: (b)(6), 02-010-01-0225 - logic femoral ps cem left sz 2.5, (b)(6), 02-012-45-2515 - lgc tibial fit tray cem sz 2.5f / 1.5t, (b)(6), 200-02-35 - three peg patella 35mm.The product associated with the reported event is within the scope of recall z-0021-2022: however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately 133 months after a left total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, pain, swelling.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.
|
| |
|
Search Alerts/Recalls
|
|
|