| Catalog Number |
242277 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during inspection at the in-house distribution center, it was observed that the c-mount endoscope, mitek lock 70 degree x 2.7 mm x 75 mm device was cloudy.This event did not occur during surgery.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: to date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: investigation summary the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: distal tip damaged distal tip fiber damaged distal cover glass/negative scratched/residue proximal cover glass residue coupler missing o-ring at back end.Broken (2+ pieces) : device fractured visual : deformed/bent per service report, this complaint was confirmed.The optical and tube were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.Based on the investigation findings, the faulty parts were identified as the root cause for the device failure during the service evaluation.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Search Alerts/Recalls
|
|
|