| Model Number |
EPIQ DIAGNOSTIC ULTRASOUND SYSTEM |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported the swivel mechanism of their epiq cvx ultrasound system¿s control panel did not lock properly while transporting the unit within the user facility.The failure occurred outside of clinical use and no patient or user was harmed as a result of the issue.The field service engineer replaced the solenoid to resolve the immediate issue.An investigation is underway to determine the root cause of the issue.Results of the investigation will be included in a follow up report upon its completion.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips service engineer replaced the solenoid to repair the system.A thorough investigation was performed to determine the cause of the reported problem.The investigation concluded a failure of the rotational solenoid prevented the release of the pin to the locked position.The ultrasound system has been returned to service with no similar issues reported.
|
| |
|