• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 4.5 HEALIX ADVANCE PEEK W/CORD; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4.5 HEALIX ADVANCE PEEK W/CORD; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 222303
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information or the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device lot number and no non-conformances related to the reported complaint condition were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event or Problem Description
It was reported that during an unspecified surgical procedure the 4.5 healix advance peek w/cord device was being used when the anchor was broken off.All broken parts were removed.Changed to another one to continue the surgery and the same problem occurred with a 2nd 4.5 healix advance peek w/cord device.Another device was used to complete the surgery.There was no delay in the procedure reported.There were no adverse patient consequences reported.No additional information was provided.This is report 2 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5 HEALIX ADVANCE PEEK W/CORD
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key20077610
Report Number1221934-2024-02980
Device Sequence Number18057821
Product Code HWC
UDI-Device Identifier10886705021390
UDI-Public(01)10886705021390(10)279L707(17)270131
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K120449
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number222303
Device Lot Number279L707
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2024
Initial Report FDA Received Date08/27/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-