• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC COOK® MULTI-USE HOLMIUM LASER FIBER; GEX LASER INSTRUMENT, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK® MULTI-USE HOLMIUM LASER FIBER; GEX LASER INSTRUMENT, SURGICAL Back to Search Results
Model Number G23668
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the 'cook multi-use holmium laser fiber' broke during a retrograde intrarenal surgery (rirs) procedure in the kidney.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Additional information regarding event details has been requested but is not available at this time.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.D4: model # = gpn #.E3: customer occupation = unknown.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation it was reported the 'cook multi-use holmium laser fiber' broke during a retrograde intrarenal surgery (rirs) procedure in the kidney.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), and quality control (qc) procedures were conducted during the investigation.Visual inspection of the returned complaint device was also conducted.One device was returned for investigation.The fiber was noted to be broken at the connection hub.The break looked charred and melted.Cook has concluded that the device was manufactured to specification.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no related non-conformances reported for lot.A complaint history database search showed no other related complaints associated with the failure mode for the complaint device lot.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.Cook also reviewed product labeling; the ifu supplied with the device states the following in consideration of the reported failure mode: - ¿warnings: ¿do not bend fiber at sharp angles.¿ ¿limitations on reprocessing: the laser fiber can be reprocessed up to 20 times, which may not reflect the actual number of uses.¿ ¿precautions: ¿do not improperly handle the fiber¿ ¿how supplied: ¿ upon removal from package, inspect the product to ensure no damage has occurred.¿ based upon the available information, inspection of the returned device, and results of the investigation, cook has concluded that the cause for the complaint could not be established.The appropriate personnel have been notified, and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK® MULTI-USE HOLMIUM LASER FIBER
Common Device Name
GEX LASER INSTRUMENT, SURGICAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key20077986
Report Number1820334-2024-01121
Device Sequence Number15348498
Product Code GEX
UDI-Device Identifier10827002236681
UDI-Public(01)10827002236681(17)260831(10)15636219
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K163197
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Model NumberG23668
Device Catalogue NumberHLF-M273-H30
Device Lot Number15636219
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/12/2024
Supplement Date Received by Manufacturer12/03/2024
Initial Report FDA Received Date08/27/2024
Supplement Report FDA Received Date12/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-