It was reported that during a t12 to pelvis posterior spinal fusion case, the tip of a probe broke off into the patient's bone and could not be removed.The tip was left in-situ, embedded in the bone.No further patient or case impact was reported.No additional information was provided.
|
|
No device was returned for evaluation and no photographs were provided for review.No operative notes and/or radiograph images were provided for review of usage/technique.A note with the reported event indicated that the patient had hard bone.Based on the information obtained, the complaint was unable to be confirmed and the root cause was unable to be determined; however, the patient's hard bone as well as any off-axis excessive force likely caused or contributed to the fracture.Labeling review: "warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient." "the instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury." "care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." "the physical characteristics required for many instruments do not permit them to be manufactured from implantable materials.If any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted.".
|