| Model Number |
OTV-S7H-1NA-10E |
| Medical Device Problem Codes |
Poor Quality Image (1408); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E2/e3: no information available for the occupation of the initial reporter or whether they are a healthcare professional.The investigation is ongoing.A supplemental report will be submitted, when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported, that the subject device presented a cut sheath.There were no reports of patient harm.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: d8, e1, h2, h3, h4, h6, h11 the device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: damage on cable unit, water tightness is lost.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: there is damage on cable unit and therefore water tightness is lost.Olympus will continue to monitor field performance for this device.
|
| |
|