| Model Number |
4554 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) and left ventricular (lv) lead exhibited fluctuating lv impedance measurements in august 2024 from a baseline of 700-900 ohms to high out-of-range pace impedance measurements greater than 3,000 ohms.This left ventricular (lv) lead remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) and left ventricular (lv) lead exhibited fluctuating lv impedance measurements in august 2024 from a baseline of 700-900 ohms to high out-of-range pace impedance measurements greater than 3,000 ohms.This left ventricular (lv) lead remains in service.No adverse patient effects were reported.Additional information was received that this lv lead was explanted and replaced due to lead fracture.No additional adverse patient effects were reported.The lead was retained by the explanting facility and was likely discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) and left ventricular (lv) lead exhibited fluctuating lv impedance measurements in (b)(6) 2024 from a baseline of 700-900 ohms to high out-of-range pace impedance measurements greater than 3,000 ohms.This left ventricular (lv) lead remains in service.No adverse patient effects were reported.Additional information was received that this lv lead was explanted and replaced due to lead fracture.No additional adverse patient effects were reported.The lead was retained by the explanting facility and was likely discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.It was reported that this left ventricular (lv) lead exhibited fluctuating lv impedance measurements in (b)(6) 2024 from a baseline of 700-900 ohms to high out-of-range pace impedance measurements greater than 3,000 ohms.Additional information was received that this lv lead was explanted and replaced due to lead fracture.No additional adverse patient effects were reported.The lead was retained by the explanting facility and was likely discarded.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of pacing impedance high out of range was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|