ALIGN TECHNOLOGY, INC. INVISALIGN SYSTEM; ALIGNER, SEQUENTIAL Back to Search Results
Model Number INVISALIGN SYSTEM - MODERATE
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Abscess (1690); Tooth Loss (4831)
Date of Event 07/01/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The current instructions for use (ifu) contains the following: "precautions - a tooth that has been previously traumatized or significantly restored may be aggravated.In rare instances, the useful life of the tooth may be reduced, the tooth may require additional dental treatment such as endodontic and/or additional restorative work, and/or the tooth may be lost".Align's clinical review of available records indicate that tooth ur1 had a poor prognosis prior the use of treatment (pre-existing root canal treatment and metal-porcelain crown with traumatic occlusion).No conclusive evidence has been provided that supports or opposes that the invisalign system aligners caused or contributed to the reported symptoms.This event is being filed as an mdr as the treating doctor reported that the patient had tooth extraction (serious injury), and the invisalign system aligners were being used.The date of event (b3) is based on the timelines reported to align and compared to the patient information.There is no conclusive evidence to confirm the date of the reported symptoms.
 
Event or Problem Description
The treating doctor reported that the patient had symptoms of severe abscess on ur1 and tooth extraction (ur1).No additional information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVISALIGN SYSTEM
Common Device Name
ALIGNER, SEQUENTIAL
Manufacturer (Section D)
ALIGN TECHNOLOGY, INC.
2820 orchard parkway
san jose CA 95134
Manufacturer (Section G)
ALIGN TECHNOLOGY, INC.
2820 orchard parkway
san jose CA 95134
Manufacturer Contact
harper shore
2820 orchard parkway
san jose, CA 95134
MDR Report Key20081234
Report Number2953749-2024-02528
Device Sequence Number7810963
Product Code NXC
UDI-Device Identifier00816063020196
UDI-Public(01)00816063020196(10)0186824569(13)240607(91)19560727XXN
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K241412
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberINVISALIGN SYSTEM - MODERATE
Device Catalogue Number9001
Device Lot Number186824569
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2024
Initial Report FDA Received Date08/27/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/07/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Disability;
Patient Age35 YR
Patient SexFemale
-
-