Model Number 1286
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, there was an issue removing the catheter.There was no problem removing it from the patient.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information: b5, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.A visual inspection of the returned photo noted that an image of the physical catheter showing clear tpe damage.It was reported that the manometry catheter was difficult to remove from patient.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, there was an issue removing the catheter as the distal part was a bit destroyed.There was no problem removing it from the patient.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MANOSCAN
Common Device Name
SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL)
Manufacturer (Section D)
GIVEN IMAGING VIETNAM CO LTD
suite 6a, 6th floor
ho chi minh city
VM 
Manufacturer (Section G)
GIVEN IMAGING VIETNAM CO LTD
suite 6a, 6th floor
ho chi minh city
VM  
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20081271
Report Number3007662958-2024-00007
Device Sequence Number12678586
Product Code FFX
UDI-Device Identifier04260167482484
UDI-Public04260167482484
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K151086
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 10/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1286
Device Catalogue Number1286
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2024
Supplement Date Received by Manufacturer10/01/2024
Initial Report FDA Received Date08/27/2024
Supplement Report FDA Received Date10/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/28/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age75 YR
Patient SexUnknown
Patient Weight80 KG
-
-