MAUDE Adverse Event Report: TORNIER S.A.S. TORNIER PERFORM ANATOMIC GLENOID CORTILOC PEGGED S35; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. TORNIER PERFORM ANATOMIC GLENOID CORTILOC PEGGED S35; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number DWE402
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Pain (1994); Numbness (2415); Muscle/Tendon Damage (4532)
Date of Event 05/09/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that in 2019, the patient received a right shoulder replacement with a tornier aequalis ascend implant and experienced pain and numbness in her shoulder, arm, and hand afterward.Although x-rays taken showed no issues, her pain worsened, limiting arm movement.In 2024, she underwent a reverse shoulder replacement after discovering that a part of the implant had come loose, damaging the rotator cuff and causing shoulder separation.The surgeon found that the loose piece was "chewed up," creating severe scar tissue, with the polyethylene glenoid reported as grossly loose, causing a central cavitary defect.The stem was retained, while other components were discarded.The patient had a left shoulder replacement in 2015 with no complications.
 
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.
 
Additional Manufacturer Narrative
Correction to fda registration number from 0001649390 - te (blooming) to 3000931034 - te (mtbonnot), d3, g1, d2a, d2b, and h6 ( method code).The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the catalog number and lot number were not communicated.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported that in 2019, the patient received a right shoulder replacement with a tornier aequalis ascend implant and experienced pain and numbness in her shoulder, arm, and hand afterward.Although x-rays taken showed no issues, her pain worsened, limiting arm movement.In 2024, she underwent a reverse shoulder replacement after discovering that a part of the implant had come loose, damaging the rotator cuff and causing shoulder separation.The surgeon found that the loose piece was "chewed up," creating severe scar tissue, with the polyethylene glenoid reported as grossly loose, causing a central cavitary defect.The stem was retained, while other components were discarded.The patient had a left shoulder replacement in 2015 with no complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER PERFORM ANATOMIC GLENOID CORTILOC PEGGED S35
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin MN 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin MN 38330
FR   38330
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key20081306
Report Number0001649390-2024-00404
Device Sequence Number7580639
Product Code KWS
UDI-Device Identifier03700434021165
UDI-Public03700434021165
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K111902
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberDWE402
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/31/2024
Supplement Date Received by Manufacturer11/12/2024
Initial Report FDA Received Date08/27/2024
Supplement Report FDA Received Date12/09/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
Patient Weight76 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-