| Catalog Number |
8881501160 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that there was an open needle inside the clinic's box of syringes and the person unpacking the box got stabbed when she reached in.She is bleeding because of it.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned samples of this event was requested but hasn't been received till now, no sample for evaluation.The dhr of this lot was reviewed without any similar issue, the retained samples were inspected and they all meet the specification.The event can't be confirmed, the root cause can't detected currrently.No action will be taken currently, this issue will be monitored.A supplemental report will be submitted if further information received.
|
| |
|