Additional information provided in d.9., h.3., h.6.And h.11.One opened probe was received, with a tip protector, in a pouch.The sample was visually inspected and found to be conforming.The sample was then functionally tested for actuation, aspiration and cut.The sample was found to be conforming for all functional tests.A review of the device history record traceable to the lot number obtained from the device¿s radio frequency identification tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned sample was found to be conforming, therefore aspiration and cut failures as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the returned probe was manufactured to specifications.All probes are hundred percent visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|