EXACTECH, INC. TRULIANT TIB IMP PS INSERT SZ 3 12MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-022-35-3012 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Loss of Range of Motion (2032); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Swelling/ Edema (4577); Insufficient Information (4580)
|
| Date of Event |
06/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately 49 months after a right total knee replacement procedure, the patient underwent a revision procedure to address loosening, debonded femoral.Revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: concomitant devices: 6433972, 02-020-11-0330 - truliant ps cem fem ps cem right sz 3.N/a, 02-012-45-3020 - lgc tibial fit tray cem sz 3f/2t.N/a, 200-07-29 - advanced patella 29m 3 peg implant.The product associated with the reported event is within the scope of recall z-0023-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and loosening or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs and relevant clinical information were not provided.D6a: corrected.H6: corrected investigation clinical codes.
|
| |
|