| Model Number |
VTICM5 13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/25/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault.In a separate visit the lens was exchanged for a replacement lens and the problem was resolved.It has been noted that the surgeon elected a different length lens than that initially suggested by the icl calculation software.Therefore, there is not enough safety and effectiveness data to support implantation in this patient category.Cause of the event is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Claim #: (b)(4).Secondary surgical intervention was performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).D10: injector and cartridge model changed g3: aware date changed to 19-aug-2024.
|
| |
|
Additional Manufacturer Narrative
|
|
A2: dob (b)(6) 1993.B5: a healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault.In a separate visit the lens was exchanged for a replacement lens and the problem was resolved.Cause of the event is unknown.D4: udi (b)(4).Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault.In a separate visit the lens was exchanged for a replacement lens and the problem was resolved.Cause of the event is unknown.
|
| |
|
Search Alerts/Recalls
|