| Model Number |
71562-01 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Abbott diabetes care (adc) received a medwatch report which reported the following information: a customer reported an adhesive issue with the adc sensor.The sensor prematurely detached and the customer was unable to obtain glucose readings.There was no third party treatment provided for the reported issue.Adc customer service attempted to contact the customer 3 times to gain additional details regarding this event; however, all follow up attempts were unsuccessful.Based on the information provided, there was no report of serious injury associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This issue does not meet reportability criteria; however, it is being reported to the fda as the complaint was received via user report.Extended investigation is pending at this time.The product has been requested back for investigation.A follow-up report will be submitted upon completion of investigation activities or if additional information is obtained.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs(device history review) for the fs libre sensor kit was reviewed and the dhrs showed the sensor kit passed all tests prior to release.If product is returned, a physical investigation will be performed per adc's established processes and procedures and a report will be submitted upon completion of investigation.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Abbott diabetes care (adc) received a medwatch report which reported the following information: a customer reported an adhesive issue with the adc sensor.The sensor prematurely detached and the customer was unable to obtain glucose readings.There was no third party treatment provided for the reported issue.Adc customer service attempted to contact the customer 3 times to gain additional details regarding this event; however, all follow up attempts were unsuccessful.Based on the information provided, there was no report of serious injury associated with this event.
|
| |
|