CARDINAL HEALTH 200, LLC SALEM SUMP; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8888266197
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the anti reflux valve snapped at the blue and white part.This was then replaced and the new one also broke at the white and blue intersection.
 
Additional Manufacturer Narrative
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured on february 20, 2024.A sample was not received for the investigation.Because a sample was not returned, we were unable to perform functional and visual evaluations to confirm the reported condition and determine the root cause.If a sample is received at a later date, the complaint will be reopened, and the investigation will be updated accordingly.A corrective action is not applicable at this time.Functional testing and visual inspections are performed according to our current quality standards and inspection procedures.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SALEM SUMP
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH 200, LLC
3651 birchwood drive
waukegan IL 60085
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
3651 birchwood drive
waukegan, IL 60085
5086183640
MDR Report Key20087674
Report Number1423537-2024-00139
Device Sequence Number7663746
Product Code KNT
UDI-Device Identifier10192253012866
UDI-Public(01)10192253012866(10)2404406264
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K851444
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8888266197
Device Catalogue Number8888266197
Device Lot Number2404406264
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2024
Initial Report FDA Received Date08/28/2024
Date Device Manufactured02/20/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-