|
The investigation determined that a lower than expected nt-probnp ii (nbnp2) result was obtained when a level 3 biorad qc fluid was processed using vitros nbnp2 lot 0260 on a vitros xt 7600 integrated system.A definitive assignable cause of the event was not established.Historical qc results leading up to the event indicate acceptable performance of vitros nbnp2 lot 0260.Additionally, ongoing tracking and trending of complaint data did not identify any signals to suggest there is a systemic quality issue with vitros nbnp2 lot 0260.An instrument issue cannot be ruled out as a contributor to the event, as no diagnostic precision testing was conducted around the time the lower than expected result was obtained to verify the performance of the vitros xt 7600 integrated system.It is possible improper handling of the biorad qcs contributed to the event, however, this could not be confirmed, as it was not possible to determine whether the customer handled the biorad level 3 qc according to the instructions prior to processing the fluid on the instrument.
|