Catalog Number 195-180
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The remainder of the investigation remains in progress.A supplemental report will be provided after completion.
 
Event or Problem Description
The consumer reported a false positive result with the binaxnow covid-19 antigen self-test performed on (b)(6) 2024 on a nasal sample.Initial testing was performed on (b)(6) 2024 via an unknown brand antigen test which generated a negative result.The consumer indicated they were experiencing symptoms such as a bad cold, sneezing, and a runny nose.The consumer confirmed there was no injury due to the test results.Additionally, the consumer confirmed there was no delay or impact in treatment.
 
Additional Manufacturer Narrative
Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 784573 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 195-180/ lot: 784573, test base part number 195-430wl/ lot: 784573.The lot met the required release specifications.A review of the complaints reported as false positive patient results (confirmed and unconfirmed, conflicting results) related to kit lot 784573 showed that the complaint rate is (b)(4).Abbott diagnostics scarborough, inc.Was unable to determine the exact root cause of the reported issue; however, it could have possibly been related to patient sample.
 
Event or Problem Description
The consumer reported a false positive result with the binaxnow covid-19 antigen self-test performed on (b)(6) 2024 on a nasal sample.Initial testing was performed on (b)(6) 2024 via an unknown brand antigen test which generated a negative result.The consumer indicated they were experiencing symptoms such as a bad cold, sneezing, and a runny nose.The consumer confirmed there was no injury due to the test results.Additionally, the consumer confirmed there was no delay or impact in treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BINAXNOW COVID-19 AG SELF TEST 4CT
Common Device Name
CORONAVIRUS ANTIGEN DETECTION TEST SYSTEM.
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
10 southgate road
scarborough ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough, ME 04074
6613888803
MDR Report Key20092831
Report Number1221359-2024-00588
Device Sequence Number18102396
Product Code QKP
UDI-Device Identifier00811877011729
UDI-Public(01)00811877011729(17)250317(10)000784573C
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EUA210264
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/17/2025
Device Catalogue Number195-180
Device Lot Number000784573C
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/16/2024
Supplement Date Received by Manufacturer09/25/2024
Initial Report FDA Received Date08/28/2024
Supplement Report FDA Received Date09/27/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/31/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexFemale
Patient Weight113 KG
Patient RaceWhite
-
-