| Catalog Number |
195-180 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The remainder of the investigation remains in progress.A supplemental report will be provided after completion.
|
| |
|
Event or Problem Description
|
|
The consumer reported a false positive result with the binaxnow covid-19 antigen self-test performed on (b)(6) 2024 on a nasal sample.Initial testing was performed on (b)(6) 2024 via an unknown brand antigen test which generated a negative result.The consumer indicated they were experiencing symptoms such as a bad cold, sneezing, and a runny nose.The consumer confirmed there was no injury due to the test results.Additionally, the consumer confirmed there was no delay or impact in treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 784573 with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 195-180/ lot: 784573, test base part number 195-430wl/ lot: 784573.The lot met the required release specifications.A review of the complaints reported as false positive patient results (confirmed and unconfirmed, conflicting results) related to kit lot 784573 showed that the complaint rate is (b)(4).Abbott diagnostics scarborough, inc.Was unable to determine the exact root cause of the reported issue; however, it could have possibly been related to patient sample.
|
| |
|
Event or Problem Description
|
|
The consumer reported a false positive result with the binaxnow covid-19 antigen self-test performed on (b)(6) 2024 on a nasal sample.Initial testing was performed on (b)(6) 2024 via an unknown brand antigen test which generated a negative result.The consumer indicated they were experiencing symptoms such as a bad cold, sneezing, and a runny nose.The consumer confirmed there was no injury due to the test results.Additionally, the consumer confirmed there was no delay or impact in treatment.
|
| |
|