| Model Number |
00711099 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure, the suctioned polyps could not be located in the cannister.User facility personnel had to sift through the cannister to locate the polyps resulting in a procedure delay.The patient procedure was completed successfully, and no injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The etrap-polyp trap was discarded and is not available for evaluation.Without the return or evaluation of the subject device a root cause could not be determined.Steris offered in-service training on the proper use and operation of the etrap-polyp trap however, the user facility declined.The instruction for use states, "inspect the package and device for shipping or handling damage, i.E.Cracked clear vessel, bent, misshapen, or missing specimen retrieval strainer(s), cracked, torn, or missing connection tube.If damage is evident do not use these devices, save them for return, and contact your local product specialist.Caution: hang trap ensuring that the suction canister's tubing does not kink.Kinked tubing will prevent the etrap from functioning properly, or may decrease or eliminate the ability to suction effectively." the device history record (dhr) was reviewed and confirmed the lot was manufactured to specification.A complaint review confirmed this event to be isolated for the subject lot.No additional issues have been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The unit subject of the reported event was discarded by the user facility and therefore not available for further evaluation.However, based on recent similar complaints, it is likely that there was a slight deformation with the bridge inside the molded vessel, such that it was slanted in a slightly downward position which caused a small gap between the tray and vessel.This gap was wide enough that small polyps were able to be suctioned over the top of the tray during the retrieval process.The bridge is designed to be slanted slightly upwards during the manufacturing process to ensure there is no gap between the tray and vessel.To date, steris has received a total of eight complaints reporting that polyps were not retained within the etrap.The complaints were received across four different lots (the other three lots were unknown as the product was discarded), manufactured between december 2023 - august 2024.The four known lots were manufactured using the same mold tool.Additional investigation of the mold tool used to produce the vessels determined that water used to cool the molds during the production process may have been inconsistently delivered due to an issue with the water line.This may have resulted in excess heat to a small number of vessels, resulting in the observed deformation.This mold tool was replaced in august 2024.An analysis of all product manufactured on this mold tool for this 9-month period determined there were a total of (b)(4) manufactured across 284 lots.This results in a reported failure rate of approximately (b)(4), further evidencing the extremely intermittent nature of the issue.The etrap is also a historically high turnover product i.E., customers order and use the product quickly.As of january 2025, nearly all of the customers who had ordered product during the 9-month time period have already reordered newer product (after august 2024); therefore, it is reasonable to assume that most or all of the product from this time period has already been used.This is further supported based on the fact that the most recent complaint for this issue was received in august 2024.Based on the overall analysis, and corrective actions already taken, further action is not planned at this time.Steris will continue to monitor under our post-market surveillance procedures to ensure the product continues to perform as intended.
|
| |
|
Search Alerts/Recalls
|
|