| Catalog Number |
8065198201 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A nurse reported that an ophthalmic suture wire was broken easily.Details of surgery and patient harm were not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
One opened suture, in a pouch, in a blister was received for the report of wire is broken easily.Sample was visually inspected and found to be nonconforming with suture is broken.A dimensional inspection for suture diameter and a functional test for suture tensile strength was performed and the sample was found to be conforming for both tests.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned sample was found to be non-conforming visually with broken suture, however the sample was conforming for all functional and dimensional testing associated with the reported event.Therefore a suture broken easily as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the suture was manufactured to specifications.Suture material samples are tested at incoming for tensile strength and diameter and visually inspected for any defects such as discoloration and frays.Sutures are also 100% inspected by trained operators for discoloration and frays and for proper attachment during the manufacturing process.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|