Catalog Number 8065198201
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
A nurse reported that an ophthalmic suture wire was broken easily.Details of surgery and patient harm were not reported.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
One opened suture, in a pouch, in a blister was received for the report of wire is broken easily.Sample was visually inspected and found to be nonconforming with suture is broken.A dimensional inspection for suture diameter and a functional test for suture tensile strength was performed and the sample was found to be conforming for both tests.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned sample was found to be non-conforming visually with broken suture, however the sample was conforming for all functional and dimensional testing associated with the reported event.Therefore a suture broken easily as described in the complaint was not confirmed and a root cause cannot be determined for the complaint as described by the customer.No action was taken as the suture was manufactured to specifications.Suture material samples are tested at incoming for tensile strength and diameter and visually inspected for any defects such as discoloration and frays.Sutures are also 100% inspected by trained operators for discoloration and frays and for proper attachment during the manufacturing process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALCON CLOSURE SYSTEM (ACS), NYLON
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20101070
Report Number2523835-2024-02017
Device Sequence Number7664478
Product Code GAR
UDI-Device Identifier00380651982017
UDI-Public(01)00380651982017(17)270531(11)220617(10)14UNH9
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K930517
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065198201
Device Lot Number14UNH9
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/01/2024
Supplement Date Received by Manufacturer03/05/2025
Initial Report FDA Received Date08/29/2024
Supplement Report FDA Received Date04/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-