| Model Number |
VICM5 12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
08/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.In a separate visit the lens was removed.In the same visit but different surgery a replacement lens of the same model and size but different power was implanted.The problem was resolved.Cause reported as a patient related factor.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: d4: primary unique device identifier (udi) #: (b)(4)."udi related data quality updates only." h3-device evaluation: lens returned in a liquid vial.Visual inspection found no visible damage to lens.Dimensional and functional inspection found the lens to be within specifications.Claim# (b)(4).
|
| |
|