| Catalog Number |
SI-09600 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "a first placement attempt is made with reference si-09600, lot 14f21l0107, inserting through the right subclavian vein.Upon inserting the temporary transvenous pacemaker, the guidewire is advanced approximately halfway; however, resistance is encountered, and the guidewire is removed, revealing it is kinked.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that "a first placement attempt is made with reference si-09600, lot 14f21l0107, inserting through the right subclavian vein.Upon inserting the temporary transvenous pacemaker, the guidewire is advanced approximately halfway; however, resistance is encountered, and the guidewire is removed, revealing it is kinked.
|
| |
|
Search Alerts/Recalls
|