Model Number OM-10000
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Tw id# (b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event or Problem Description
The hospital reported during a coronary artery bypass procedure, the acrobat-i stabilizer would not tighten and was stiff.There was a 5 minute procedural delay when we had to run and get a new stabilizer.A replacement device was used to complete the procedure.There were no patient harm or injury reported.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Trackwise # (b)(4).Corrected sections: d4--unique identifier (udi) # corrected from "(b)(4).The investigation has been started since the complaint was received, the following contents have been conducted: 1.Dhr review: the dhr of the reported lot has been reviewed.No non-conformity relevant with the reported issue is observed.All the products had been performed 100% mechanical function test during production.They all passed the function test which demonstrate the knob can be tightened and can also tighten the flexlink arm.2.Trend review: in the last 12 months, the occurrence rate is found to be 0.058% which is under anticipated occurrence rate.3.Device was not received for evaluation, the specific root cause for this failure could not be determined.4.Review complaint historical data, the failure knob difficult/ unable to tighten-arm does not lock happened before and has been investigated.The most probable root cause for the knob tighten abnormal could be acme nut lead in thread broken for the reason that rotating to lose the knob anti-clockwise rapidly, or rotating the knob leaner or too much strength used, which caused acme screw misaligned with the acme nut lead in thread, acme nut lead in thread broken, then the knob could not be tighten, and link arm could not be tightened.There were no ncmrs, rework, or deviations documented for the reported batch, based on above review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT-I STABILIZER
Common Device Name
STABILIZER, HEART
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key20107593
Report Number2242352-2024-00994
Device Sequence Number8340796
Product Code MWS
UDI-Device Identifier00607567700581
UDI-Public(01)00607567700581(10)3000357109(17)251124
Combination Product (Y/N)N
Initial Reporter StateFL
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/24/2025
Device Model NumberOM-10000
Device Catalogue NumberOM-10000
Device Lot Number3000357109
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2024
Supplement Date Received by Manufacturer10/21/2024
Initial Report FDA Received Date08/30/2024
Supplement Report FDA Received Date11/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.; UNKNOWN.
Patient Age57 YR
Patient SexMale
Patient Weight110 KG
-
-