• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRO MED INSTRUMENTS GMBH DORO® QR3 SKULL CLAMP TEFLON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRO MED INSTRUMENTS GMBH DORO® QR3 SKULL CLAMP TEFLON Back to Search Results
Model Number 3003-009
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 08/12/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Customer informed us on (b)(6) 2024 that one of our products was involved in a case (posterior cervical, supine position) in which the patient sustained a laceration during repositioning from supine to prone.
 
Additional Manufacturer Narrative
It is not suspected that the slight play in the arch holder of the rocker arm contributed to the reported slippage.No further deviations in the product could be detected.From our experience, the pinning technique is a major factor in regards to the occurence of slippages.Therefore, we suspect that the pinning of the patient's head may not have been optimal as described in the instruction manual: "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® QR3 SKULL CLAMP TEFLON
Common Device Name
DORO® QR3 SKULL CLAMP TEFLON
Manufacturer (Section D)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
nathalie ockenfuss
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
MDR Report Key20114094
Report Number3003923584-2024-00029
Device Sequence Number15278101
Product Code HBL
UDI-Device Identifier04250435500781
UDI-Public04250435500781
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K001808
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Repair,Inspection
Type of Report Initial
Report Date (Section B) 08/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number3003-009
Device Catalogue Number3003-009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2024
Initial Report FDA Received Date08/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-