| Model Number |
CSL-TR105-44 |
| Medical Device Problem Code |
Failure to Seal (4070)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
08/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on august 12 a gastric sleeve procedure was performed with a coolseal trinity.During the procedure, the doctor did 5 or 6 cuts along the stomach line and tried to seal the vessel which was under 7mm, the device was not sealing and the doctor encountered 2ml of bleeding on at least two occasions.The doctor swapped the device in less than two minutes and was able to control the bleeding.No additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
| |
|
Search Alerts/Recalls
|