Model Number CSL-TR105-44
Medical Device Problem Code Failure to Seal (4070)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 08/12/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on august 12 a gastric sleeve procedure was performed with a coolseal trinity.During the procedure, the doctor did 5 or 6 cuts along the stomach line and tried to seal the vessel which was under 7mm, the device was not sealing and the doctor encountered 2ml of bleeding on at least two occasions.The doctor swapped the device in less than two minutes and was able to control the bleeding.No additional information available.
 
Additional Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOLSEAL¿ TRINITY
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOLDER SURGICAL
331 south 104 street
ste 200
louisville CO 80027
Manufacturer (Section G)
BOLDER SURGICAL
331 south 104 street
ste 200
louisville CO 80027
Manufacturer Contact
daniel guevara
562 parkway
coyol free park zone b24
alajuela 20102
CR   20102
MDR Report Key20118372
Report Number3010377594-2024-00008
Device Sequence Number17884646
Product Code GEI
UDI-Device Identifier00850346007061
UDI-Public(01)00850346007061(17)260128(11)240129(10)24A29RW
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K211579
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/28/2026
Device Model NumberCSL-TR105-44
Device Catalogue NumberCSL-TR105-44
Device Lot Number24A29RW
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2024
Initial Report FDA Received Date09/02/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-