| Model Number |
U1054260B |
| Medical Device Problem Codes |
Gas Output Problem (1266); Decrease in Pressure (1490)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A trilogy100 ventilator was returned to the manufacturer service center for foam replacement per field action: c&r 2021-05/06.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer service center, the device powered on and operated as it should.The inlet air path assembly and removable air path foam was replaced per remediation.In addition, the pm label and enclosure feet were replaced.The device was returned to the technician for further testing.The device failed a test step for o2 low flow.The device's active exhalation control module needs to be replaced to address the issue.No repairs performed.The device was scrapped.
|
| |
|