• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number U1054260B
Medical Device Problem Codes Gas Output Problem (1266); Decrease in Pressure (1490)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
A trilogy100 ventilator was returned to the manufacturer service center for foam replacement per field action: c&r 2021-05/06.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer service center, the device powered on and operated as it should.The inlet air path assembly and removable air path foam was replaced per remediation.In addition, the pm label and enclosure feet were replaced.The device was returned to the technician for further testing.The device failed a test step for o2 low flow.The device's active exhalation control module needs to be replaced to address the issue.No repairs performed.The device was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key20123563
Report Number2518422-2024-55673
Device Sequence Number7482523
Product Code CBK
UDI-Device Identifier00606959028944
UDI-Public(01)00606959028944
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberU1054260B
Device Catalogue NumberU1054260B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2024
Initial Report FDA Received Date09/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/25/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-